Model XE

Instrument Applications:
- Leak Testing under pressure (max. 50 psig)
- Leak Rates of 1cc/min or 0.5cc/min (optional) and higher
- Small to medium size parts (1,000cc or less typical)
Leak Test Simulator: Use our Leak Test Simulator to optimize your application and define your micro-flow sensor/instrument.
Features:
- Simple to use
- Automated Pressure Test Circuit
- Two Line, Alpha Numeric Display
- Digital I/O
- Serial interface
- Leak-Tek© Program
- Expansion tank, externally mounted
- Calibrated Orifice (optional)
- Reliable Instrument
- Tamper-Proof
- Temperature stable - Does not require zero'ing or daily calibration
- Direct Flow measurement is not volume dependent
- Measurement is Direct, Not Calculated - Reducing Errors
Specifications
Dimensions:
- 15″W, 4.75″H,10.5″D (Does not include connectors, fitting, and accessories)
- Expansion tank and pressure regulators are externally mounted
Gases:
- Dry, clean gases: air, nitrogen
- For other gases, consult ATC, Inc.
Flow Measurement Ranges:
- 8, 15, 40, 80, and 200cc/min. - Measurement Uncertainty: +/- 3% of reading, Calibration Range
Pressure Measurement:
- 0-50 psig - Measurement uncertainty: +/- 1% of full scale
- In-Line Filter (Internal) - Included