Model IPE

Instrument Applications:
- 1 or 2 of the following tests:
- Pressure Leak Test from 0.03 cc/min to 500 psig
- Constriction or Blockage Flow Test, up to 5% part diameter constriction
- Mass Extraction - vacuum tests from 0.02 cc/min at 2 psia
- 1 or 2 independent test channels
- Testing up to two set of parts simultaneously
- Sort function - testing multiple parts and sorting the rejects. Applicable as the IGLS measurement is not volume sensitive!
- Large volume parts, complex parts and short cycle time
Leak Test Simulator: Use our Leak Test Simulator to optimize your application and define your micro-flow sensor/instrument.
Available Options: Please see ATC's options chart.
Features:
- Micro-Flow Sensor - The Intelligent Gas Leak Sensor (IGLS)
- Single or Dual Test Channels with one or two of the following:
- Leak Test Channel Pressure Testing: Automated test with large volume Quick-Fill features
- Constriction / Blockage Flow Channel: with Intelligent Gas Flow Sensor (IGFS) and unique pressure control
- Vacuum Test Channel: Mass Extraction tests to 2 psia vacuum
- Optional Sort Test Controller: Leak Test multiple parts, and sorting out the rejected parts.
- For Each Test Channel:
- Built in Equivalent Diameter (ED) and valve.
- Character Display Interface.
- Digital and analog I/O Interface
- Serial Interface for each test channel.
Applicable Micro-Flow Sensors (IGLS):
- Model IL2-C & Model IL2-L
- Flow / Constriction / Blockage Tests: Model IF2-HF
- Vacuum Leak Test: Model IL2-M
The IPE interfaces the IGLS or the IMFS, which provides the source for measurement and control, for your application. Special vacuum components and valves are used.
Each leak test instrument is set and tested with customer parts (if practical), prior to shipping from ATC, Inc.
Specifications
Dimensions & Access:
- 24”W, 23”H, 12”D (Does not include connectors & fittings)
- Dual access doors, NEMA 12 instrument enclosure
- Expansion tank and pressure regulators are externally mounted
Gases:
- Dry, clean gases: air, nitrogen
- For other gases, consult ATC, Inc.
- Digital Inputs: 5 VDC, NPN Opto-Isolated for start, stop, type, pressure
switch, verify Digital Outputs: 30 VDC-20 mA, NPN, opto-isolated for
pass, fail, clamp, test type, exhaust and custom
Analog outputs: single channel, 0-5 VDC pressure control - Pneumatic connections: Test Ports
Air supply: ¼” to ¾” Swagelok
(Application Dependent) - RS232 Serial Interface
- Power Supply:
115VAC-60Hz
220VAC-50Hz (optional)
Print Model IPE - PDF (465K)