Please Note: You are viewing the unstyled version of ATC, Inc. Either your browser does not support CSS (Cascading Style Sheets) or it is disabled. As a result, much of this website will not look the way it was intended, although all of its contents will be accessible to you. For more information, visit our Browser Support page.


Skip to Primary Site Navigation, Secondary Site Navigation, Content, Company Address


Model IL2-M

Model IL2-M

Instrument Applications:

  • 2 psia to 24.5 psia (mild vacuum to low pressure)
  • Leak Rates: 1 x 10 -5 sccs at 2 psia
  • Used for Mass Extraction applications
  • Can be used for applications with low positive pressure and tight leak specification
  • Detection of 2 micron defects of larger
Leak Test Simulator button

Features:

  • Micro-Flow Sensor - The Intelligent Gas Leak Sensor (IGLS)
  • Measures volume flow - using ATC's patented ALF design
  • Measures absolute downstream pressure and temperature
  • Stainless steel rugged construction
  • No moving parts
  • No over-flow sensitivity
  • Measurement units select by user, typically mm3/min (µL/min)
  • Lowest capable measurement, 1x10-6 sccs at 2 psia

Flow Range:

IF2-M Models
  • 0-25 mm3 /min (micro liters/min)
  • 0-50 mm3 /min (micro liters/min)
  • 0-100 mm3 /min (micro liters/min)
  • 0-250 mm3 /min (micro liters/min)
  • 0-500 mm3 /min (micro liters/min)
  • 0-990 mm3 /min (micro liters/min)

Sensor Type: Accelerated Laminar Flow (ALF).

Measurement Uncertainty: ±2% of reading (1% optional), calibrated range.

Pressure Range:

  • 0-15; 30; (psi-absolute)
  • Type: Absolute, silicon micro-machined
  • Measurement Uncertainty: 0.2% of Full Scale (FS)
  • Maximum over Pressure: 1.2 times full range/scale

Temperature Range:

  • Operating and Calibrated: 10 to 45°C
  • Storage: -25 to 50°C
  • Sensor Type: RTD 100 Ohms
  • Measurement uncertainty at calibrated range: 0.5°C

Resolution: 16 bits A/D and 16 bits D/A

Interface:

  • Serial port
  • Digital I/O: Start/Stop, type clamp and more…
  • Analog I/O
  • Power: 115 or 220 VAC, single phase
  • Optional: 2-line character display

Max. Differential Pressure: 10 psi, differential

Gases Used:

  • Use dry non-condensing and clean gases
  • Air, Nitrogen, Helium, Argon, & Carbon Dioxide
  • Other Gases Available, consult ATC, Inc.

Response Time:

  • Step function: from no flow to full range - 50 ms or less
  • Sensor only, no volume
Mass Extraction Tests

Click to Enlarge

Print Model IL2-M - PDF (427K)