Please Note: You are viewing the unstyled version of ATC, Inc. Either your browser does not support CSS (Cascading Style Sheets) or it is disabled. As a result, much of this website will not look the way it was intended, although all of its contents will be accessible to you. For more information, visit our Browser Support page.


Skip to Primary Site Navigation, Secondary Site Navigation, Content, Company Address


Model IL2-HP

Model IL2-HP

Instrument Applications:

  • High Pressure applications
  • Range: 500 - 2200 psig
  • Leak Rates: 0.05-2 cc/min
Leak Test Simulator button

Features:

  • Micro-Flow Sensor - The Intelligent Gas Leak Sensor (IGLS)
  • Measures volume flow - using ATC's patented ALF design
  • Stainless steel rugged construction
  • No moving parts
  • No over-flow sensitivity
  • Max. Differential Pressure: Up to 500 psi DP

Flow Range:

IL2-M Model
  • 0-1 cc/min (mL/min)
  • 0-3 cc/min (mL/min)
  • 0-5 cc/min (mL/min)
  • 0-10 cc/min (mL/min)
  • 0-25 cc/min (mL/min)

Sensor Type: Accelerated Laminar Flow (ALF).

Measurement Uncertainty: ±2% of reading, calibrated range.

Pressure Range:

  • 0-500; 1000; 2000; (psi-absolute)
  • Type: Absolute, micro-machined
  • Measurement Uncertainty: 0.3% of Full Scale (FS)
  • Maximum over Pressure: 1.2 times full range/scale

Temperature Range:

  • Operating and Calibrated: 10 to 45°C
  • Storage: -25 to 50°C
  • Sensor Type: RTD 100 Ohms
  • Measurement uncertainty at calibrated range: 0.5°C

Resolution: 16 bits A/D and 16 bits D/A

Interface:

  • Serial port
  • Digital I/O: Start/Stop, type clamp and more…
  • Analog I/O
  • Power: 115 or 220 VAC, single phase
  • Optional: 2-line character display

Max. Differential Pressure: 500 psid (DP) or max. range of pressure sensor, whichever is smaller.

Gases Used:

  • Use dry non-condensing and clean gases
  • Air, Nitrogen, Helium, Argon, & Carbon Dioxide
  • Other Gases Available, consult ATC, Inc.

Response Time:

  • Step function: from no flow to full range - 50 ms or less
  • (Open outlet, dependent on sensor range)
Leak Signature - High Pressure

Click to Enlarge

Print Model IL2-HP - PDF (442K)