Model IL2-HP

Instrument Applications:
- High Pressure applications
- Range: 500 - 2200 psig
- Leak Rates: 0.05-2 cc/min
Features:
- Micro-Flow Sensor - The Intelligent Gas Leak Sensor (IGLS)
- Measures volume flow - using ATC's patented ALF design
- Stainless steel rugged construction
- No moving parts
- No over-flow sensitivity
- Max. Differential Pressure: Up to 500 psi DP
Flow Range:

- 0-1 cc/min (mL/min)
- 0-3 cc/min (mL/min)
- 0-5 cc/min (mL/min)
- 0-10 cc/min (mL/min)
- 0-25 cc/min (mL/min)
Sensor Type: Accelerated Laminar Flow (ALF).
Measurement Uncertainty: ±2% of reading, calibrated range.
Pressure Range:
- 0-500; 1000; 2000; (psi-absolute)
- Type: Absolute, micro-machined
- Measurement Uncertainty: 0.3% of Full Scale (FS)
- Maximum over Pressure: 1.2 times full range/scale
Temperature Range:
- Operating and Calibrated: 10 to 45°C
- Storage: -25 to 50°C
- Sensor Type: RTD 100 Ohms
- Measurement uncertainty at calibrated range: 0.5°C
Resolution: 16 bits A/D and 16 bits D/A
Interface:
- Serial port
- Digital I/O: Start/Stop, type clamp and more…
- Analog I/O
- Power: 115 or 220 VAC, single phase
- Optional: 2-line character display
Max. Differential Pressure: 500 psid (DP) or max. range of pressure sensor, whichever is smaller.
Gases Used:
- Use dry non-condensing and clean gases
- Air, Nitrogen, Helium, Argon, & Carbon Dioxide
- Other Gases Available, consult ATC, Inc.
Response Time:
- Step function: from no flow to full range - 50 ms or less
- (Open outlet, dependent on sensor range)
Print Model IL2-HP - PDF (442K)