Please Note: You are viewing the unstyled version of ATC, Inc. Either your browser does not support CSS (Cascading Style Sheets) or it is disabled. As a result, much of this website will not look the way it was intended, although all of its contents will be accessible to you. For more information, visit our Browser Support page.


Skip to Primary Site Navigation, Secondary Site Navigation, Content, Company Address


Home > Leak Testing / Detection > Micro-Flow Sensors > Model IL2-C / Model IL2-L

Model IL2-C / Model IL2-L

Model IL2-C / Model IL2-L

Instrument Applications:

  • Pressure leak testing
  • Leak Rates: 0.03 cc/min to 25 LPM
Leak Test Simulator button

Features:

  • Micro-Flow Sensor - The Intelligent Gas Leak Sensor (IGLS)
  • Measures volume flow - using ATC's patented ALF design
  • Measures absolute downstream pressure and temperature
  • Stainless steel rugged construction
  • No moving parts
  • No over-flow sensitivity
  • Max. Differential Pressure: Up to 500 psi.

Flow Ranges:

  • 0-1 cc/min
  • 0-3 cc/min
  • 0-5 cc/min
  • 0-10 cc/min
  • 0-25 cc/min
  • 0-50 cc/min
  • 0-100 cc/min
  • 0-250 cc/min
  • 0-500 cc/min

Sensor Type: Accelerated Laminar Flow (ALF).

Measurement Uncertainty: ±1% (0.6% optional) of reading, calibrated range.

Pressure Range:

  • 0-15; 30; 100; 250 (psi-absolute)
  • Type: Absolute, micro-machined
  • Measurement Uncertainty: 0.2% of Full Scale (FS)
  • Maximum over Pressure: 1.2 times full range/scale

Temperature Range:

  • Operating and Calibrated: 10 to 45°C
  • Storage: -25 to 50°C
  • Sensor Type: RTD 100 Ohms
  • Measurement uncertainty at calibrated range: 0.5°C

Resolution: 16 bits A/D and 16 bits D/A

Interface:

  • Serial port
  • Digital I/O: Start/Stop, type clamp and more…
  • Analog I/O
  • Power: 115 or 220 VAC, single phase
  • Optional: 2-line character display

Max. Differential Pressure: 500 psid (DP) or max. range of pressure sensor, whichever is smaller.

Gases Used:

  • Use dry non-condensing and clean gases
  • Air, Nitrogen, Helium, Argon, & Carbon Dioxide
  • Other Gases Available, consult ATC, Inc.

Response Time:

  • Step function: from no flow to full range - 50 ms or less
  • Sensor only, no volume
Leak Signature

Click to Enlarge

Print Model IL2-C / IL2-L - PDF (440K)


Top of Page