Please Note: You are viewing the unstyled version of ATC, Inc. Either your browser does not support CSS (Cascading Style Sheets) or it is disabled. As a result, much of this website will not look the way it was intended, although all of its contents will be accessible to you. For more information, visit our Browser Support page.


Skip to Primary Site Navigation, Secondary Site Navigation, Content, Company Address


Home > Leak Testing / Detection > Micro-Flow Sensors > Model IF2-C / Model IF2-L

Model IF2-C / Model IF2-L

Model IF2-C

Instrument Applications:

  • Medium and low flow range measurement
  • Portable transfer standard
  • Calibration of Leak Test orifices and instruments
  • Measurement of gas mixtures
  • Pressure up to 500 PSIG
Leak Test Simulator button

Features:

  • Micro-Flow Sensor - The Intelligent Gas Flow Sensor (IGFS)
  • Measures volume flow - using ATC's patented ALF design
  • Dual range calibration, automatic range switches
  • Measures absolute downstream pressure and temperature
  • Optional: Compensation for various gas properties, temperature and viscosity
  • Stainless steel rugged construction
  • No moving parts
  • No over-flow sensitivity
  • Max. Differential Pressure: Up to 500 psi

Flow Ranges:

Handheld IGFS
  • 0-1 mL/min
  • 0-3 mL/min
  • 0-5 mL/min
  • 0-10 mL/min
  • 0-25 mL/min
  • 0-50 mL/min
  • 0-100 mL/min
  • 0-250 mL/min
  • 0-500 mL/min

Sensor Type: Accelerated Laminar Flow (ALF).

Measurement Uncertainty: ±1% (0.5% optional) of reading, volumetric flow, calibrated.

Pressure Range:

  • 0-15; 30; 100; 250 (psi-absolute)
  • Type: Absolute, micro-machined
  • Measurement Uncertainty: 0.2% of Full Scale (FS)
  • Maximum over Pressure: 1.2 times full range/scale

Temperature Range:

IF2-C & IF2-L
  • Operating and Calibrated: 10 to 45°C (Higher temperature available, consult ATC, Inc.)
  • Storage: -25 to 50°C
  • Sensor Type: RTD 100 Ohms
  • Measurement uncertainty at calibrated range: 0.5°C

Resolution: 16 bits A/D and 16 bits D/A

Interface:

  • Serial port
  • Digital I/O: Start/Stop, type clamp and more…
  • Analog I/O
  • Power: 115 or 220 VAC, single phase
  • Optional: 2-line character display

Max. Differential Pressure: 500 psid (DP) or max. range of pressure sensor, whichever is smaller.

Gases Used:

  • Use dry non-condensing and clean gases
  • Air, Nitrogen, Helium, Argon, & Carbon Dioxide, Hydrogen, Natural Gas, Moisturized Gases
  • Other Gases Available, consult ATC, Inc.

Response Time:

  • Step function: from no flow to full range - 50 ms or less
  • Sensor only, no volume