Please Note: You are viewing the unstyled version of ATC, Inc. Either your browser does not support CSS (Cascading Style Sheets) or it is disabled. As a result, much of this website will not look the way it was intended, although all of its contents will be accessible to you. For more information, visit our Browser Support page.


Skip to Primary Site Navigation, Secondary Site Navigation, Content, Company Address


Model E2

Model E2

Instrument Applications:

  • Leak testing under pressure (max. 160 psig)
  • Leak rates of 0.1 cc/min and higher
  • Medium sized parts, higher throughput
  • Cleanroom or industrial applications
  • User friendly operator interface - stand alone applications
Leak Test Simulator button

Leak Test Simulator: Use our Leak Test Simulator to optimize your application and define your micro-flow sensor/instrument.

Available Options: Please see ATC's options chart.

Features:

  • Micro-Flow Sensor - The Intelligent Gas Leak Sensor (IGLS)
  • Automated pressure test circuit with Balance and Quick-Fill features
  • Front TFT graphical color display with touch screen
  • Built in Equivalent Diameter (ED)
  • Digital and analog I/O Interface
  • Ethernet or serial interface
  • Stainless steel enclosure for cleanroom
  • Multiple test profiles

Applicable Micro-Flow Sensors (IGLS):

Specifications

Dimensions:

  • 12”W, 12”H, 12”D (Does not include connectors & fittings)
  • Expansion tank and pressure regulators are externally mounted

Gases:

E2 Specifications

Click to Enlarge

  1. Power Supply:
    115VAC-60Hz
    220VAC-50Hz (optional)
  2. Pneumatic connections:
    Test Ports
    Air supply: ¼” Swagelok
  3. Digital Inputs: 5 VDC, NPN Opto-Isolated for start, stop, type, pressure switch, verify
    Digital Outputs: 30 VDC-20 mA, NPN,
    Opto-isolated for pass, fail, clamp, test type, exhaust and custom
    Analog outputs: single channel, 0-5 VDC pressure control
  4. User selected
    RJ45 Ethernet or
    RS232 Serial Interface

Print Model E2 - PDF (762K)